广州医药 ›› 2021, Vol. 52 ›› Issue (5): 35-39.DOI: 10.3969/j.issn.1000-8535.2021.05.007

• 论著 • 上一篇    下一篇

高通量透析模式对改善维持性血液透析患者β2微球蛋白清除率及高血压的影响

何同林1, 何兰1, 刘俊2, 刘俊华3, 阮宪东1, 朱艳红1   

  1. 1 广东省广州市增城区新塘医院(广州 511340);
    2 南方医科大学南方医院(广州 510515);
    3 南方医科大学南方医院增城分院(广州 511340)
  • 收稿日期:2021-01-29 出版日期:2021-09-20 发布日期:2021-11-24
  • 基金资助:
    广东省医学科研基金项目(C2018039)

The effect of high flux dialysis mode on the improvement of β2-microglobulin clearance and hypertension in maintenance hemodialysis patients

HE Tonglin1, HE Lan1, LIU Jun2, LIU Junhua3, RUAN Xiandong1, ZHU Yanhong1   

  1. 1 Xintang Hospital, Zengcheng District, Guangzhou City, Guangzhou 511340, China;
    2 Southern Hospital of Southern Medical University,Guangzhou 510515, China;
    3 Zengcheng Branch of Southern Hospital, Southern Medical University,Guangzhou 511340, China
  • Received:2021-01-29 Online:2021-09-20 Published:2021-11-24

摘要: 目的 探讨高通量透析模式对改善维持性血液透析(MHD)患者β2微球蛋白清除率及高血压的影响。方法 选择2018年9月1日—2018年10月31日期间在广州市增城区新塘医院血液净化中心接受MHD患者40例为实验对象(除外因自身耐受因素长期使用低通量透析患者)。采用随机数表法分两组,每组20例。对照组接受低通量血液透析,观察组接受高通量血液透析。比较两组治疗效果,并记录0个月、3个月、6个月、12个月、18个月治疗前后β2微球蛋白(β2-MG)、全段甲状旁腺激素(iPTH)、心率、血压变化,并对两组心血管疾病情况进行统计。结果 观察组总有效率为95.00%,高于对照组的20.00%(P<0.05);治疗后,观察组β2-MG、iPTH均降低,而3个月后对照组略有升高,观察组[(12.48±2.10)mg/L、(210.13±18.12)ng/L]低于对照组[(30.21±2.37)mg/L、(289.41±17.02)ng/L],观察组治疗3个月、6个月、12个月、18个月的β2-MG清除率高于对照组(P<0.05);与治疗前比较,两组的平均动脉压、收缩压均降低,且观察组低于对照组(P<0.05);对照组各时间点24小时舒张压及收缩压差异无统计学意义(P>0.05);与治疗前比较,观察组24小时舒张压、收缩压治疗后1个月、3个月、6个月、12个月、18个月均降低,且低于对照组(P<0.05);两组患者心血管疾病发生率比较:观察组有1例因急性心肌梗死住院行PCI术;对照组有2例因心力衰竭住院治疗,1例因急性冠脉综合症住院治疗。两组并发症发生率比较P>0.05。结论 高通流量透析模式可有效提高MHD患者的治疗效果,有效清除血液毒素,降低患者的高血压,减少并发症,可在临床推广使用。

关键词: 维持性血液透析, 高通量透析模式, β2微球蛋白清除率, 高血压

Abstract: Objective To investigate the effect of high flux dialysis mode on the improvement of β2-microglobulin clearance and hypertension in maintenance hemodialysis (MHD) patients. Methods Patients who received MHD at the Blood Purification Center of Xintang Hospital, Zengcheng District, Guangzhou from September 1, 2018 to October 31, 2018 (except for long-term low-flux dialysis patients due to self-tolerance factors)were selected. Random number table method was used to divide the patients into two groups,20 patients in each group. The control group received low flux hemodialysis, and the observation group received high flux hemodialysis. The treatment effects of the two groups were compared, levels of β2-microglobulin (β2-MG), intact parathyroid hormone (iPTH), changes in heart rate and blood pressure, and statistics on cardiovascular disease in the two groups before and after treatment at 3, 6, 12, and 18 months were recorded. Results The total effective rate in the observation group was 95.00%, which was higher than the 20.00% in the control group (P<0.05). After treatment, the levels of β2-MG and iPTH of the two groups decreased, and the observation group [(13.02±2.10) mg/L, (210.13±18.12) ng/L] was lower than the control group [(19.78±2.37) mg/L, (289.41±17.02)ng/L]; the β2-MG clearance rate of the observation group was higher than that of the control group at 3 months, 6 months, 12 months, and 18 months (P<0.05).Compared with that before treatment, the average arterial pressure and systolic blood pressure of the two groups were reduced, and the observation group was lower than that of the control group (P<0.05); the differences between diastolic blood pressure and systolic blood pressure at each time point in the control group were not significant (P>0.05); the 24-hour diastolic blood pressure and systolic blood pressure of the observation group decreased after 1, 3, 6, 12 and 18 months, compared with those before treatment (P<0.05), and were lower than those in the control group (P<0.05).Comparison of the incidence of cardiovascular disease between the two groups: 1 case in the observation group was hospitalized for acute myocardial infarction; 2 cases in the control group were hospitalized for heart failure and 1 case was hospitalized for acute coronary syndrome; complications incidence in the two groups were compared (P>0.05). Conclusion The high flux dialysis mode can effectively improve the treatment outcomes of MHD patients, effectively remove blood toxins, lower blood pressure, and reduce the incidence of complications. It can be used in clinical practise.

Key words: maintenance hemodialysis, high flux dialysis mode, β2-microglobulin clearance rate, hypertension